In this work, a new methodology for the dynamic calibration of a sensorized helmet is proposed. The helmet is a medical device for head motion detection, based on gyroscopes and linear accelerometers. It has been designed for monitoring head motion in patients undergoing neuro rehabilitation.
While a few solutions are available in the literature for the calibration of this particular type of helmet, no solution currently exists that allows for a complete dynamic calibration by taking into account actual head movements.
With the aim of obtaining such a dynamic calibration solution, in this work an instrumented test bench has been designed that replicates typical head movements. Ranges and frequencies for each specific head movement type in the frontal, sagittal and median plane have been retrieved from the literature; data is related to both patients and healthy subjects. Two dynamic calibration prototypes have been produced as the result of subsequent design iterations: the first is capable of reproducing only a few types of principal motions, while the second and final one allows for a complete reproduction of all the principal head movement types. The prototypes have been subjected to experimental testing; the results are analyzed and discussed in the thesis.

Abstract (italiano)

Il presente lavoro di tesi ha avuto per oggetto lo sviluppo di una nuova metodologia di taratura dinamica di caschi strumentati, utilizzati in campo neuro riabilitativo per rilevare i moti del capo dei pazienti. È stato realizzato un sistema in grado di riprodurre i movimenti tipici della testa.
Il lavoro è iniziato con una ricerca bibliografica per definire i campi di variabilità dei parametri di taratura, ovvero le rotazioni tipiche del capo sul piano frontale, sagittale e trasversale, nonché i loro campi di frequenza. In letteratura si trovano diverse configurazioni per la sensorizzazione del caschetto, con accelerometri e giroscopi, e diversi sistemi ad oggi utilizzati per la loro taratura. Ma non esistono sistemi sviluppati ad hoc, come questo, per poter fare una taratura dinamica, riproducendo proprio le condizioni operative del casco strumentato.
Sono stati realizzati e testati due prototipi, il secondo dei quali, il più complesso, realizza tutte le rotazioni del capo definite dall’analisi bibliografica.

Statistiche Download - Aggiungi a RefWorks
Tipo di EPrint:Tesi di dottorato
Relatore:Rossi, Gianluca
Dottorato (corsi e scuole):Ciclo 22 > Scuole per il 22simo ciclo > SCIENZE TECNOLOGIE E MISURE SPAZIALI > MISURE MECCANICHE PER L'INGEGNERIA
Data di deposito della tesi:NON SPECIFICATO
Anno di Pubblicazione:20 Gennaio 2010
Parole chiave (italiano / inglese):taratura dinamica casco strumentato
Settori scientifico-disciplinari MIUR:Area 09 - Ingegneria industriale e dell'informazione > ING-IND/12 Misure meccaniche e termiche
Struttura di riferimento:Centri > Centro Interdipartimentale di ricerca di Studi e attività  spaziali "G. Colombo" (CISAS)
Codice ID:2347
Depositato il:28 Ott 2010 16:34
Simple Metadata
Full Metadata
EndNote Format

Bibliografia

I riferimenti della bibliografia possono essere cercati con Cerca la citazione di AIRE, copiando il titolo dell'articolo (o del libro) e la rivista (se presente) nei campi appositi di "Cerca la Citazione di AIRE".
Le url contenute in alcuni riferimenti sono raggiungibili cliccando sul link alla fine della citazione (Vai!) e tramite Google (Ricerca con Google). Il risultato dipende dalla formattazione della citazione.

[1] T. Pozzo, A. Berthoz, L. Lefort , Head kinematic during various motor tasks in humans - Progress in Brain Research 1989; 80: 377-383 Cerca con Google

[2] T. Pozzo, A. Berthoz, L. Lefort, Head stabilization during various locomotor tasks in humans. (Normal Subjects). Experimental Brain Research 1990; 82: 97-106 Cerca con Google

[3] T. Pozzo, A. Berthoz, L. Lefort, E. Vitte , Head stabilization during various locomotor tasks in humans. (Pazient with bilateral peripheral vestibular deficits). Experimental Brain Research 1991; 83: 208-217 Cerca con Google

[4] RL Cromwell, RA Newton, G Forrest, Influence of vision on head stabilization strategies in older adults during walking - Journal of Gerontology: Medical Sciences 2002; 57A: M442-M448 Cerca con Google

[5] CARPANETO J., MICERA S., GALARDI G., MICHELI A., CARBONCINI M.C., ROSSI B., Cerca con Google

and DARIO P., 2004, A protocol for the assessment of 3D movements of the head in persons with cervical dystonia - Clinical Biomechanics; 19: 659–663 Cerca con Google

[6] STRIMPAKOS N., SAKELLARI V., GIOFTSOS G., KAPRELI E., and OLDHAM J., 2006, Cerca con Google

Cervical joint position sense: an intra- and inter-examiner reliability study - Gait and Posture; Cerca con Google

23(1):22-31 Cerca con Google

[7] D. Cattaneo, PT, M. Ferrarin, ENG, W. Frasson, A. Casiraghi, MD, Head Control: Volitional aspects of Rehabilitation Training in Patients With Multiple Sclerosis Compared With Healthy Subjects, Arch Phys Med Rehabil Vol. 86, July 2005 Cerca con Google

[8] D. Giansanti, V.Macellari, Is it Feasible to Reconstruct Body Segment 3-D Position and Orientation Using Accelerometric Data, IEEE Transactions on biomedical engineering, Vol 50,no. 4, April 2003. Cerca con Google

[9] G. Baselli, G. Legnani, P. Franco, F. Brognoli, A. Marras, F. Quaranta, B. Zappa, Assessment of inertial and gravitational input sto the vestibular system, Journal of Biomechanics 34 (2001) 821-826. Cerca con Google

[10] P. Cappa, L. Masia, F. Patanè, Numerical Validation of Linear Accelerometer Systems for the Measurement of Head Kinematics, Journal of Biomechanical Engineering, Vol. 127, November 2005. Cerca con Google

[11] P.Cappa, F. Patanè, S. Rossi, A Redundant Accelerometric Cluster for the Measurement of Translational and Angular Acceleration and Angular Velocità of the Head, Journal of Medical Devices, March 2007. Cerca con Google

[12] P. Cappa, F. Patanè, S. Rossi, Two calibration procedures for a gyroscope-free inertial measurement system based on a double-pendulum apparatus, MEASUREMENT SCIENCE AND TECHNOLOGY, 2 April 2008. Cerca con Google

[13] C. Cattaneo, M. Ferrarin, Head Control: Voluntary Mechanism and Rehabilitation, Arch Phys Rehabil Vol. 86, July 2005. Cerca con Google

[14] Raj S.Sodhi, Evaluation of head and neck motion with the hemispherical shell method, International Journal of Industrial Ergonomics, October 1999 Cerca con Google

[15] WTFong, S K Ong and A Y C Nee, Methods for in-field user calibration of an inertial measurement unit without external equipment, MEASUREMENT SCIENCE AND TECHNOLOGY, July 2008 Cerca con Google

[16] S. Rossi, Tesi di Dottorato: Sviluppo ed utilizzo in ambito biomedico di un casco strumentato per il rilievo della cinematica della testa - Università degli Studi di Padova, 2008 Cerca con Google

[17] Y.Higashi, M.Sekimoto, F.Horiuchi, T.Kodama ,T.Yuji, T. Fujimoto, M. Sekine, T. Tamura, Monitoring rehabilitation training for hemiplegic patients by using a tri-axial accelerometer. 23rd Annual International Conference of the IEEE Engineering in Medicine and Biology Society, 25-28 October, Istanbul, Turkey Cerca con Google

[18] G. Wu, Head movement during sudden base traslation as a measure of risks for falls in the Cerca con Google

Elderly - Clinical Biomechanics; 16: 199-206, 2001. Cerca con Google

[19] HB. Menz, SR. Lord, RC. Fitzpatrick, Acceleration patterns of the head and pelvis when Cerca con Google

walking on level and irregular surfaces, Gait&Posture; 18: 35-46; 2003. Cerca con Google

[20] HB. Menz, SR. Lord, RC. Fitzpatrick, Acceleration patterns of the head and pelvis when Cerca con Google

walking are associated with risk of falling in community-dwelling older people, Journal of Gerontology; 58A, 5: 446-452; 2003. Cerca con Google

[21] J.R. Morris, Accelerometry-A technique for the measurement of human body movements, Journal of Biomechanics; 6: 729–736, 1973. Cerca con Google

[22] A.J. Padgaonkar, K.W. Krieger, A.I. king, Measurement of angular acceleration of a rigid body using linear accelerometers - Journal of Applied Mechanics; 42: 552–558, 1975. Cerca con Google

[23] G.S. Nusholtz, Geometric methods in determining rigid-body dynamics, Experimental Mechanics, 33 (2), pp.153-158, 1993. Cerca con Google

[24] E. O. Deobelin, Measurement System Application and Design, Ed. Mc Graw-Hill. Cerca con Google

[25] P. Brandi, A. Salvadori, Modelli matematici elementari, Ed. Bruno Mondadori. Cerca con Google

[26] V. Pirola, Cinesiologia:il movimento umano applicato alla rieducazione e alle attività sportive, Edi-ermes. Cerca con Google

Download statistics

Solo per lo Staff dell Archivio: Modifica questo record